MEDICAL AND TECHNICAL REQUIREMENTS FOR MEDICAL MONITORS FOR MAMMOGRAPHY RESEARCH

https://doi.org/10.22328/2079-5343-2020-11-2-119-122

Abstract

Purpose: determination of technical characteristics when choosing medical viewing monitors, affecting the quality of the viewed image.

Materials and methods: Analysis of current regulatory documents, domestic and world scientific  literature, regulatory parameters for medical viewing monitors.

Results and discussions: In Russian and foreign publications, there are few articles on the assessment and analysis of medical monitors, and there are no regulatory documents. According to the technical documentation of medical viewing monitors presented in Russia, the basic requirements for medical  viewing monitors that are used when working on modern digital mammography systems have been  formed. The article analyzes the main parameters of monitors that affect the quality of the visualized image, also highlights the main parameters of digital detectors installed on mammographs and gives recommendations for choosing a medical monitor that meets the parameters of the mammograph. The main parameters for monitors are highlighted, such as resolution, brightness to contrast ratio, response time, which, if properly correlated, significantly improve the quality of work of a radiologist.

Conclusions. Medical viewing monitors should meet the physico-technical parameters of the installed mammograph; to improve the quality of the work of radiologists involved in the diagnosis of mammary glands within the framework of one health care facility, it is advisable to choose medical viewing  monitors of one manufacturer; maintenance of medical viewing monitors should be performed at least twice a year.

About the Authors

O. O. Manuylova
City Clinical Hospital named after V. M. Buyanov; Central Research Institute of Radiation Diagnostics
Russian Federation
Moscow



A. Yu. Vasil’ev
Moscow State University of Medicine and Dentistry named after A. I. Evdokimov; Central Research Institute of Radiation Diagnostics
Russian Federation
Moscow



T. V. Pavlova
City Clinical Hospital named after V. M. Buyanov
Russian Federation
Moscow



References

1. Pavlova T.V., Vasil’ev A.Yu., Manuylova O.O. The Impact of Compliance with the Rules of Mammography Laying on the Timely Diagnosis of Breast Cancer (The Clinical Example). Diagnostic and intervention radiology, 2019, No. 2 (13), рр. 60–65 (In Russ.). DOI: 10.25512/DIR.2019.13.2.07.

2. Pavlova T.V., Vasil’ev A.Yu., Manuylova O.O. Method of Сone-Вeam Breast Computed Tomography (Literature Review). Radiologiya — praktika, 2019, No. 1 (73), рр. 21–27 (In Russ.).

3. Manuylova O.O., Vasil’ev A.Yu., Pavlova T.V. Medical and technical requirements for x-ray mammographs for various purposes // Radiologiya — praktika, 2018, No. 5 (71), рр. 24–30 (In Russ.).

4. Awan O, Safdar N.M., Siddiqui K.M., Moffitt R., Siegel E.L. Detection of cervical spine fracture on computed radiography images: a monitor resolution study // Acad. Radiol. 2011. Vol. 18. Р. 353–358.

5. Ekpo U.E., McEntee M.F. An evaluation of performance characteristics of primary display devices // J. Digit Imaging. 2015. doi: 10.1007/s10278-0159831-3.

6. Kimpe T., Xthona A. Quantification of detection probability of microcalcifications at increased display luminance levels. 2012; doi: 10.1007/978-3-642-31271-7_63.

7. Marchessoux C., de Paepe L., Vanovermeire O., Albani L. Clinical evaluation of a medical high dynamic range display // Med. Phys. 2016. Vol. 43. Р. 4023–4031.


Review

For citations:


Manuylova O.O., Vasil’ev A.Yu., Pavlova T.V. MEDICAL AND TECHNICAL REQUIREMENTS FOR MEDICAL MONITORS FOR MAMMOGRAPHY RESEARCH. Diagnostic radiology and radiotherapy. 2020;11(2):119-122. (In Russ.) https://doi.org/10.22328/2079-5343-2020-11-2-119-122

Views: 683


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 2079-5343 (Print)